Multimeter
17B+ DMMs
Your job requires that you have a rugged, reliable and accurate digital multimeter. The Fluke 17B+ digital multimeter does all that and more. Common basic electrical measurements plus temperature, frequency and duty cycle. Easy to use with one hand, even with gloves on, the Fluke 17B+ offers everything you need.
Your job requires that you have a rugged, reliable and accurate digital multimeter. The Fluke 17B+ digital multimeter does all that and more. Common basic electrical measurements plus temperature, frequency and duty cycle. Easy to use with one hand, even with gloves on, the Fluke 17B+ offers everything you need.
- 600 V Cat III safety rating
- 50% bigger display with bright white backlight
- Over-voltage indicator
- Frequency and temperature measurement
- Voltage, resistance, continuity, capacitance
- Input terminal for ac and dc current measurements to 10 A current
- Diode test, data hold
Accuracy Specifications
Accuracy is specified for 1 year after calibration, at operating temperatures of 18 Ā°C to 28 Ā°C, relative humidity at 0 % to 75 %. Accuracy specifications take the form of: Ā±(% of Reading + Number of Least Significant Digits).
Function | Range | Resolution | Accuracy |
AC volts (40 Hz to 500Hz)1 |
4.000 V 40.00 V 400.0 V 1000 V |
0.001 V 0.01 V 0.1 V 1 V |
1.0 % + 3 |
DC volts | 4.000 V 40.00 V 400.0 V 1000 V |
0.001 V 0.01 V 0.1 V 1 V |
0.5 % + 3 |
AC millivolts | 400.0 mV | 0.1 mV | 3.0 % + 3 |
DC millivolts | 400.0 mV | 0.1 mV | 1.0 % + 10 |
Diode test2 | 2.000 V | 0.001 V | 10% |
Resistance (Ohms) |
400.0 Ī© 4.000 kĪ© 40.00 kĪ© 400.0 kĪ© 4.000 MĪ© 40.00 MĪ© |
0.1 Ī© 0.001 kĪ© 0.01 kĪ© 0.1 kĪ© 0.001 MĪ© 0.01 MĪ© |
0.5 % + 3 0.5 % + 2 0.5 % + 2 0.5 % + 2 0.5 % + 2 1.5% + 3 |
Capacitance3 | 40.00 nF 400.0 nF 4.000 Ī¼F 40.00 Ī¼F 400.0 Ī¼F 1000 Ī¼F |
0.01 nF 0.1 nF 0.001 Ī¼F 0.01 Ī¼F 0.1 Ī¼F 1 Ī¼F |
2 % + 5 2 % + 5 5 % + 5 5 % + 5 5 % + 5 5 % + 5 |
Frequency1 Hz (10 Hz ā 100 kHz) |
50.00 Hz 500.0 Hz 5.000 kHz 50.00 kHz 100.0 kHz |
0.01 Hz 0.1 Hz 0.001 kHz 0.01 kHz 0.1 kHz |
0.1 % + 3 |
Duty Cycle1 | 1 % to 99 % | 0.1 % | 1 % typical 4 |
AC current Ī¼A (40 Hz to 400 Hz) |
400.0 Ī¼A 4000 Ī¼A |
0.1 Ī¼A 1 Ī¼A |
1.5 % + 3 |
AC current mA (40 Hz to 400 Hz) |
40.00 mA 400.0 mA |
0.01 mA 0.1 mA |
1.5 % + 3 |
AC current A (40 Hz to 400 Hz) |
4.000 A 10.00 A |
0.001 A 0.01 A |
1.5 % + 3 |
DC current Ī¼A | 400.0 Ī¼A 4000 Ī¼A |
0.1 Ī¼A 1 Ī¼A |
1.5 % + 3 |
DC current mA | 40.00 mA 400.0 mA |
0.01 mA 0.1 mA |
1.5 % + 3 |
DC current A | 4.000 A 10.00 A |
0.001 A 0.01 A |
1.5 % + 3 |
Temperature | 50 Ā°C- 400 Ā°C 0 Ā°C- 50 Ā°C -55 Ā°C- 0 Ā°C |
0.1C | 2 % Ā± 1 Ā°C Ā± 2 Ā°C 9 % Ā± 2 Ā°C |
Backlight | – | – | Yes |
1All ac, Hz, and duty cycle are specified from 1 % to 100 % of range. Inputs below 1 % of range are not specified. 2Typically, open circuit test voltage is 2.0 V and short circuit current is <0.6 mA. 3Specifications do not include errors due to test lead capacitance and capacitance floor (may be up to 1.5 nF in the 40 nF range). 4Typical means when the frequency is at 50 Hz or 60 Hz and the duty cycle is between 10 % and 90 %. |
Function | Overload Protection | Input Impedance (Nominal) | Common Mode Rejection Ratio | Normal Mode Rejection Ratio |
AC volts | 1000 V 1 | >10 MĪ© <100 pF | >60 dB at dc, 50 Hz or 60 Hz |
– |
AC millivolts | 400 mV | >1MĪ©, <100 pF | >80 dB at 50 Hz or 60 Hz | – |
DC volts | 1000 V 1 | >10 MĪ© <100 pF | >100 dB at dc, 50 Hz or 60 Hz |
>60 dB at 50 Hz or 60 Hz |
DC millivolts | 400 mV | >1MĪ©, <100 pF | >80 dB at 50 Hz or 60 Hz | – |
1 106 V Hz Max |
General Specifications | |
Maximum voltage between any terminal and earth ground | 1000 V |
Display (LCD) | 4000 counts, updates 3/sec |
Battery type | 2 AA, NEDA 15A, IEC LR6 |
Battery life | 500 hours minimum (50 hours in LED Test mode without load. The hours with load depends on the type of LED under test.) |
Temperature | |
Operating | 0 Ā°C to 40 Ā°C |
Storage | -30 Ā°C to 60 Ā°C |
Relative Humidity | |
Operating humidity | Non-condensing (<10Ā°C) ā¤90% RH at 10 Ā°C to 30 Ā°C; ā¤75% RH at 30 Ā°C to 40 Ā°C |
40 MĪ© range ā¤80% RH at 10 Ā°C to 30 Ā°C; ā¤70% RH at 30 Ā°C to 40 Ā°C |
|
Altitude | |
Operating | 2000 m |
Storage | 12,000 m |
Temperature coefficient | 0.1 X (specified accuracy) /Ā°C (<18 Ā°C or >28 Ā°C) |
Fuse protection for current inputs | 440 mA, 1000 V Fast Fuse, Fluke specified part only. 11A, 1000V Fast Fuse, Fluke specified part only. |
Size (HxWxL) | 183 x 91 x 49.5 mm |
Weight | 455 g |
IP rating | IP 40 |
Safety | IEC 61010-1, IEC61010-2-030 CAT III 600 V, CAT II 1000 V, Pollution Degree 2 |
Electromagnetic environment | IEC 61326-1: Portable |
Electromagnetic compatibility | Applies to use in Korea only |
Class A equipment (industrial broadcasting & communication equipment)1 1 This product meets requirements for industrial (Class A) electromagnetic wave equipment and seller or user should take notice of it. This equipment is intended for use in business environments and is not to be used in homes. |
Cases and Holsters | |
TPAK | Fluke TPAK ToolPakTM Magnetic Meter Hanger Kit includes:
|
Test Leads | |
TL-175 | TL175 TwistGuardā¢ Test Leads |
You must be logged in to post a review.
Reviews